| Model Number |
LC105 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been made available for evaluation.Should the device become available, or further information be provided, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Alleges consumer was lifting out of chair when it reclined all the way on its own and got stuck.Alleges consumer had to call fire department to get out of chair.
|
| |
|
Event or Problem Description
|
|
Alleges consumer was lifting out of chair when it reclined all the way on its own and got stuck.Alleges consumer had to call fire department to get out of chair.
|
| |
|
Additional Manufacturer Narrative
|
|
Only the hand control was returned.Functional and electrical testing revealed the db2 subminiature switch on the hand control's pcb that controls the recline function failed.It is unknown why this component failed.
|
| |
|
Search Alerts/Recalls
|