• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN TRAPEASE FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN TRAPEASE FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466P306X
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Stenosis (2263)
Event Date 04/19/2021
Event Type  Injury  
Manufacturer Narrative
As reported, a patient underwent placement of a trapease inferior vena cava (ivc) filter. The indication for filter insertion has not been provided. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to physical and emotional damages from stenosis and the resultant symptoms. The filter remains implanted thus unavailable for analysis. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed the trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Stenosis of the ivc is associated with all ivc filter products and does not represent a device malfunction. A protective inferior vena cava (ivc) filter may later be incorporated into a chronic post-thrombotic ilio-caval obstruction (occlusive, requiring recanalization, or nonocclusive). Obstruction of varying types of ivc filters may occur due to primary thrombosis of the filter or capture of large emboli. Permanent ivc filters have been reported to obstruct in up to 20% of patients. Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics. Without procedural films or images for review the reported event(s) could not be confirmed. Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported in the legal brief, a patient underwent placement of a trapease vena cava filter. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to physical and emotional damages from stenosis and the resultant symptoms. As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain, suffering and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN TRAPEASE
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014-2802
7863138372
MDR Report Key11761523
MDR Text Key248760659
Report Number1016427-2021-04957
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number466P306X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/22/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/03/2021 Patient Sequence Number: 1
-
-