Model Number 400SMTXSFT0410
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2021-00924.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils) and a microcatheter.During the procedure, while advancing a smart coil through the microcatheter, the physician experienced resistance and the smart coil would not advance any further; therefore, the smart coil was removed.While attempting to advance another smart coil within its introducer sheath, resistance was encountered, and subsequently, the coil would not advance past its initial position within its introducer sheath; therefore, it was also removed.The procedure was completed using new smart coils.There was no adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned smart coil revealed offset coil winds near the proximal end of its embolization coil.If the introducer sheath is not properly aligned within the hub of the microcatheter prior to advancement, resistance may be encounter and damage such as offset coil winds may occur.During functional testing, the smart coil was able to advance through a demonstration microcatheter with resistance due to the kinks in the pusher assembly.The kinks in the pusher assembly were likely incidental to the complaint and may have occurred during packaging for return to penumbra.Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced during advancement, and the device may not be advanced out of the introducer sheath.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.Then the smart coil was able to advance through its introducer sheath without an issue.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-00924.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11766171
Report Number3005168196-2021-00925
Device Sequence Number19969693
Product Code HCG
UDI-Device Identifier00814548015408
UDI-Public00814548015408
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/19/2025
Device Model Number400SMTXSFT0410
Device Catalogue Number400SMTXSFT0410
Device Lot NumberF96410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/05/2021
Supplement Date Received by Manufacturer05/20/2021
Initial Report FDA Received Date05/03/2021
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Date Report Sent to FDA01/01/2005
-
-