• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH SCREWDRIVER BIT AXSOS T15, AO FITTING 4.0MM LOCKING SET; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH SCREWDRIVER BIT AXSOS T15, AO FITTING 4.0MM LOCKING SET; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 702753
Medical Device Problem Codes Break (1069); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event or Problem Description
It was reported there was difficulty inserting the screw due to artificial bone obstruction.The user attempted to insert the screw by power driver, but the screwdriver bit broke.
 
Additional Manufacturer Narrative
D9 / h3 was updated to indicate the device was available for return/returned.H6 method, results, conclusion code corrected based on device evaluation.The reported event could be confirmed, since the device was returned for evaluation and matches the reported failure mode.Device inspection revealed the following: the received screwdriver bit shows signs of normal and prolonged usage however its tip is found to be broken.The tip was not received for evaluation.The breakage surface shows signs of slight rotational deformation/waves.Absence of plastic deformation and presence of an uneven breakage indicates that the breakage was instantaneous due to high torsional and bending overload.However, it cannot be excluded that residual stress from previous usage may have also contributed in the breakage.The average hardness of the drill was observed to be 61.1 hrc as required against a range of 58.0 hrc ¿ 62.0 hrc.The returned screwdriver bit was found to be within specifications.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation, the root cause of the failure is user related.The breakage of the tip happened due to a combination of torsional and bending overload during the surgery.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
It was reported there was difficulty inserting the screw due to artificial bone obstruction.The user attempted to insert the screw by power driver, but the screwdriver bit broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREWDRIVER BIT AXSOS T15, AO FITTING 4.0MM LOCKING SET
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11767592
Report Number0008031020-2021-00206
Device Sequence Number11823340
Product Code LXH
UDI-Device Identifier07613327005028
UDI-Public07613327005028
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number702753
Device Catalogue Number702753
Device Lot NumberJ03730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Initial Date Received by Manufacturer 04/07/2021
Supplement Date Received by Manufacturer06/23/2021
Initial Report FDA Received Date05/03/2021
Supplement Report FDA Received Date06/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-