| Catalog Number |
9030375 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device has not been returned.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported the driver was being used with a 10nm torque handle and that the tip of the driver broke off in the head of the screw and was cold welded to the screw head.The tip was not able to be retrieved from the patient.The surgeon proceeded to fold the tab over with the punch on the ulnar cap.No additional patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Note corrections to d9/h3 device was returned to manufacturer.The reported event could be confirmed, since the device was returned and found to be in the condition reported.The device inspection revealed the following: visual inspection: the received hex driver shows signs of intense usage.The drill was found to be broken at the hex tip.The remaining edges of the hex tip are blunt and appear to be slightly twisted.The breakage surface reveals a smooth surface.This confirms a sudden brittle fracture.All these signs are evident enough to suggest that there was an intense usage of the driver including high torsional and bending load which led to a forced fracture.Functional inspection: due to the nature of the returned device, a functional inspection was not possible.As per the dimensional inspection and hardness testing, the returned driver was found to be within specifications.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to misuse of the device / failure to follow the surgical technique.If any further information is provided, the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported the driver was being used with a 10nm torque handle and that the tip of the driver broke off in the head of the screw and was cold welded to the screw head.The tip was not able to be retrieved from the patient.The surgeon proceeded to fold the tab over with the punch on the ulnar cap.No additional patient complications were reported.
|
| |
|