| Model Number |
6532 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 15mm x 3.25mm nc quantum apex balloon catheter was advanced for dilation, but failed to cross into the stent.However, upon removal, the device came apart into two separate pieces.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The device was microscopically and visually examined.There were numerous kinks to the device.There was a separation of the hypotube 53.7cm from the strain relief.The separated ends of the hypotube were ovaled, indicating the hypotube was kinked prior to separation.There was contrast and blood in the inflation lumen.There was contrast in the loosely folded balloon.The device also had tip damage.The damage to the device is consistent to damage that can occur during preparation, use during the procedure, or post procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 15mm x 3.25mm nc quantum apex balloon catheter was advanced for dilation, but failed to cross into the stent.However, upon removal, the device came apart into two separate pieces.No further information available.
|
| |
|