• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number CX01A
Medical Device Problem Codes Clumping in Device or Device Ingredient (1095); Coagulation in Device or Device Ingredient (1096)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider via a manufacturer representative regarding a patient who was implanted with a spinal product type for unknown indication.It was reported that while performing a kyphoplasty procedure and preparing for cement for injection, it was reported that shortly after pouring in the monomer and trying to mix ,the cement became very "clumpy." staff continued to try to mix the cement but it became increasingly difficult as they cement continued to harden in a very short period of time (less than 30 seconds).As a result cement and mixer was discarded and it was replaced with a different cx01b cement and mixer.The case was then successfully complete without issue.Delay was only a couple of minutes to open and mix replacement cement (less than 5 minutes).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHX HV-R BONE CEMENT
Common Device Name
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key11768509
Report Number1030489-2021-00561
Device Sequence Number3104654
Product Code NDN
UDI-Device Identifier00643169097803
UDI-Public00643169097803
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K102397
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberCX01A
Device Catalogue NumberCX01A
Device Lot NumberEL70211
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2021
Initial Report FDA Received Date05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-