| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Will be returned.
|
| |
|
Event or Problem Description
|
|
It was reported the comprehensive stem size primary stem packaging was broken, the plastic was cracked through to the inside packing containing the stem.Attempts have been made and there is no further information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Cmp-(b)(4).Reported event was confirmed as visual inspection of the returned product found the outer box has some damage.The outer blister is cracked.The inner blister is not damaged.The sterility, or breach thereof, cannot be determined as the sterile packaging was returned opened.Device history record was reviewed and no discrepancies relevant to the reported event were found.The condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|