MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC RETRACTOR BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC RETRACTOR BODY Back to Search Results
Model Number 03.816.001
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Reporter is a j&j sales representative.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, during testing of a set at loaners department, the retractor body items component parts broke off in pieces.There was no patient involvement.This complaint involves one (1) device.This report is for (1) retractor body.This report is 1 of 1 (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part: 03.816.001, lot: 9450467, manufacturing site: hägendorf, release to warehouse date: 03.Aug.2015.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the retractor body (p/n: 03.816.001, lot number: 9450467) was received at us customer quality (cq).Visual inspection of the complaint device showed the two screw angulation blades had broken.Device failure/defect identified? yes.Dimensional inspection: a dimensional inspection was not performed due to post-manufacturing damage.Document/specification review: current and manufactured were reviewed.No design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the two screw angulation blades had broken.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETRACTOR BODY
Common Device Name
RETRACTOR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11769916
Report Number2939274-2021-02210
Device Sequence Number14397483
Product Code GAD
UDI-Device Identifier10705034730622
UDI-Public(01)10705034730622
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.816.001
Device Catalogue Number03.816.001
Device Lot Number9450467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Initial Date Received by Manufacturer 04/08/2021
Supplement Date Received by Manufacturer05/05/2021
Initial Report FDA Received Date05/04/2021
Supplement Report FDA Received Date06/01/2021
Patient Sequence Number1
-
-