• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IDEAL IMPLANT IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IDEAL IMPLANT IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT Back to Search Results
Model Number 63501
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Alleged deflation.
 
Event or Problem Description
Elective removal of implant.
 
Additional Manufacturer Narrative
No manufacturing defect was found on macroscopic or microscopic examination of the explanted device.Elective removal was reported as the reason for explantation.
 
Event or Problem Description
Deflation resulting in explantation.
 
Additional Manufacturer Narrative
No manufacturing defect or other abnormality was found on macroscopic or microscopic examination of the explanted device.No leak could be reproduced during explant analysis.The cause could not be determined.Update/correction to sections b1, b2, b4, b5, d6b, d9, e1, g6, h2, h3, h6, and h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDEAL IMPLANT STRUCTURED BREAST IMPLANT
Common Device Name
SALINE-FILLED BREAST IMPLANT
Manufacturer (Section D)
IDEAL IMPLANT
14881 quorum drive
suite 925
dallas TX 75254
MDR Report Key11771793
Report Number3011491947-2021-00107
Device Sequence Number14251067
Product Code FWM
UDI-Device Identifier10851795006121
UDI-Public(01)10851795006121(241)63501(17)181113
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/13/2018
Device Model Number63501
Device Catalogue Number635
Device Lot Number61482
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer04/27/2021
04/27/2021
Initial Report FDA Received Date05/04/2021
Supplement Report FDA Received Date06/18/2021
07/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-