| Model Number |
GF-UE160-AL5 |
| Medical Device Problem Codes |
Connection Problem (2900); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.Inspection of the device found dents on the c-cap probe unit.Image check on the um (ultrasound image) was not able to be performed due to damaged m-zif lock pin.Peeling on cement of the objective lens were observed and low tensions on control knob were determined.Investigation is going therefore the cause of the reported phenomenon is unknown at this time.This report will be supplemented accordingly following investigations.
|
| |
|
Event or Problem Description
|
|
Scope connection issue was reported during reprocessing.There was no patient involvement, no user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to inform that upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.The following sections were updated: g3, g6, h2 and h10.
|
| |
|