STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HAR9F
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was detected.The blade, jaw, teflon pad appear to be intact.A gap near the min buttons was observed.The min and max buttons were pressed and released multiple times.It was observed the min button would remain in the down position when released.Compared to min button in off position, the min button would stick.There was no issues observed with the max button.The device was connected to a g11 scalpel generator (id# 0766, software rev 2014_1) using the appropriate hand piece (serial id# (b)(4)).The scalpel was tested and passed the initial testing.While pressing and releasing the min button it was observed that the device continued to activate.Min button was observed to stick during the functional testing.While the min button was stuck, the generator alerted both visually and audibly to indicate that the device was still activating.A review of the dhr supports that the device met all inspections and test criteria prior to release from stryker.Therefore the most likely root cause of the reported event is handle halves separation due to improper connection of instrument to handpiece (improper torque wrench application technique used) or handling conditions subsequent to distribution from stryker.The instructions for use (ifu) state: do not torque the instrument by hand or damage may occur to the hand piece.Do not use any means other than the torque wrench to attach or detach the instrument from the hand piece.Take care to avoid damage to the shears when removing the torque wrench from the instrument.Use the torque wrench to tighten the instrument onto the hand piece.Turn the wrench clockwise while holding the hand piece until it clicks twice, indicating that sufficient torque has been applied to secure the instrument.To ensure properly assembly, do not grip the instrument handle while applying torque with the torque wrench.Caution: do not torque the instrument by hand or damage may occur to the hand piece.Do not use any means other than the torque wrench to attach or detach the instrument from the hand piece.Remove the torque wrench from the instrument.Do not discard the disposable torque wrench until the completion of the surgical case.The torque wrench is used for removal of the instrument from the hand piece following the procedure.In the event the torque wrench falls out of the sterile field, replace with a sterile torque wrench.Do not re-sterilize the disposable torque wrench.Caution: take care to avoid damage to the shears when removing the torque wrench from the instrument.While holding the hand piece, loosen the instrument by turning the torque wrench counterclockwise.Continue to loosen by turning the instrument manually to completely unscrew it from the hand piece.To avoid user or patient injury in the event that accidental activation occurs, the instrument blade, clamp arm, and distal end of the shaft should not be in contact with the patient, drapes, or flammable materials while not in use.During and following activation in tissue, the instrument blade and clamp arm may become hot.Avoid unintended contact with tissue, drapes, surgical gowns, or other unintended sites at all times.Incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade failure, and should be avoided.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this happens, there may be a system failure signaled by a continuous tone or alert screen when either of the foot pedals or hand control buttons is depressed.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the harmonic scalpel kept firing and never turned off, continuous activation.The event occurred at the beginning of the case on the first try.The doctor removed the device and replaced it.There was no patient injury or medical intervention and minimal extended procedure time was reported to be a few minutes only to replace the device.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
A review of mdrs filed since february 20th, 2020, through august 5th, 2022, was conducted.Since february 20th, 2020, there have been no additional reports of serious injury or surgical intervention to preclude serious injury or permanent impairment related to this specific failure mode.Based on the analysis of mdr data compared to the firm¿s sales figures, the likelihood of this specific failure mode causing or contributing to another serious injury should the malfunction occur is considered remote.As such, we will no longer be reporting for the failure mode of ultrasonic scalpels (us) min activation button sticking resulting in continued activation.
 
Event or Problem Description
It was reported that the harmonic scalpel kept firing and never turned off, continuous activation.The event occurred at the beginning of the case on the first try.The doctor removed the device and replaced it.There was no patient injury or medical intervention and minimal extended procedure time was reported to be a few minutes only to replace the device.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key11772759
Report Number0001056128-2021-00025
Device Sequence Number11823450
Product Code NLQ
UDI-Device Identifier07613327348248
UDI-Public07613327348248
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K170456
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/08/2023
Device Model NumberHAR9F
Device Catalogue NumberHAR9FRR
Device Lot Number12393995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/09/2021
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date05/04/2021
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-