Model Number PFSR110006
Medical Device Problem Codes Break (1069); Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a navio tka procedure, the insertion of an anspach cutter l-5b into the long attachment revealed a resistance that could hardly be overcome and revealed a spiral-shaped material, there was possibly metal debris.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
H6: the navio long attachment, pfsr110006 intended for use in treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was not confirmed.The bur passed easily through the long attachment lumen.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints associated with the part identified prior events.A relationship between the reported event and the device could not be established as there was no damage or evidence of an obstruction found with the device during the visual and functional inspection.Although the reported problem was not confirmed a contributing factor may be disintegration of the internal bearings causing an obstruction in the long attachment.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO LONG ATTACHMENT
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11773470
Report Number3010266064-2021-00325
Device Sequence Number19916115
Product Code OLO
UDI-Device Identifier00885556628607
UDI-Public00885556628607
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K160537
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR110006
Device Catalogue NumberPFSR110006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer07/15/2021
Initial Report FDA Received Date05/04/2021
Supplement Report FDA Received Date07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-