SMITH & NEPHEW MEDICAL LTD. RENASYS GO; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801496
Device Problems False Alarm (1013); Defective Alarm (1014); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  malfunction  
Event Description
It was reported that during npwt with renasys go in a sacral wound patient a false blockage alarm activates.Later the dressing had a leak and the device did no activate the corresponding alarm.No injury or other complications were reported.The treatment finished with the same device.
 
Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.Inappropriate device alarms and alarms that occur in the absence of a pump malfunction will not directly cause or contribute to serious injury or death, even in a clinical setting.Additional conditions must also exist (e.G.Surgical / treatment conditions; patient factors; the unavailability of back-up or alternate therapies for greater than 24 hours) before an injury could possibly occur, and even under those conditions, the risk of patient harm is low.This event is considered not reportable pursuant to 21 cfr part 803.
 
Event Description
It was reported that during npwt with renasys go in a sacral wound patient a false blockage alarm activates.Later had a leak and the device did no activate the corresponding alarm.No injury or other complications were reported.The treatment finished with a competitor device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS GO
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11773609
MDR Text Key249013940
Report Number8043484-2021-00999
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00040565126944
UDI-Public00040565126944
Combination Product (y/n)N
PMA/PMN Number
K112127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66801496
Device Catalogue Number66021496
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/12/2021
Initial Date FDA Received05/04/2021
Supplement Dates Manufacturer Received04/12/2021
08/13/2021
Supplement Dates FDA Received05/05/2021
08/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-