• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 10CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 10CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000041
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up or inspection the opsite flexifix 10cmx10m ctn 1 was creased.The procedure was completed using the same device.No case involved.
 
Additional Manufacturer Narrative
H3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review found a small number of similar instances of the reported event.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was not used for treatment and was returned for analysis.The roll was creased and was coming away from the carrier paper at one side, thus establishing a relationship between the event and the device.It has not been possible to identify a definitive root cause.Probable root cause is damaged caused during the packing of the roll within the carton during manufacture.Quality checks are in place to prevent damaged goods leaving the production site and so this may be as a result of a missed quality check.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX 10CMX10M CTN 1
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11773627
Report Number8043484-2021-01004
Device Sequence Number14514242
Product Code KGX
UDI-Device Identifier05000223411473
UDI-Public5000223411473
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2023
Device Catalogue Number66000041
Device Lot Number2037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer10/13/2021
Initial Report FDA Received Date05/04/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-