BOSTON SCIENTIFIC CORPORATION ALTRUA 60; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
|
Back to Search Results |
|
| Model Number |
S601 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker had one year remaining longevity a year ago.During the recent check, the device was still at one year remaining.Boston scientific technical services (ts) discussed that the device has been implanted for ten years.Elective replacement indicator (eri) was latched.The patient will be seen in 90 days.To this date, this device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.The product has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker had one year remaining longevity a year ago.During the recent check, the device was still at one year remaining.Boston scientific technical services (ts) discussed that the device has been implanted for ten years.Elective replacement indicator (eri) was latched.The patient will be seen in 90 days.To this date, this device remains in service.No adverse patient effects were reported.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.The product has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.Correction to field e.Initial reporter.Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was eri.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared eri earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker had one year remaining longevity a year ago.During the recent check, the device was still at one year remaining.Boston scientific technical services (ts) discussed that the device has been implanted for ten years.Elective replacement indicator (eri) was latched.The patient will be seen in 90 days.To this date, this device remains in service.No adverse patient effects were reported.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|