| Model Number |
VENUM14100 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.(expiry date: 08/2022).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement, the device allegedly failed to expand.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo venous stent system 9f which was product catalog number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not returned for evaluation.X-ray image provided demonstrating an hour-glass like deformation of the stent inside the vessel in the area of the stent brake confirmed that the stent adhered to the inner catheter stent brake during deployment which leads to a confirmed investigation result for expansion issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 08/2022), g3.H11: h6 (method and result).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement, the device allegedly failed to expand.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|