• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH SENSIS VIBE HEMO; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH SENSIS VIBE HEMO; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 11007641
Medical Device Problem Code Patient Data Problem (3197)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the sensis vibe hemo system.The user reported that patient data from a previous case was showing in the most recent case on the patient explorer.At this time, we are unaware of any impact to the state of health of any patient or user involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.The detailed investigation showed that the problem was not an error in the generation of x-ray images, but a problem in the creation of reports, which only takes place after the examination.This could be proven by examining the log files.Furthermore, it does not affect all studies in general, but the error only occurs when two studies of the same patient have been made.In this case, both studies of the same patient are linked, because they have the same sequence number field (segno field).A possible mixing of data of different patients is not possible.The workaround step is to manually update the study segno field for this study record to the next value so that the study link to the study data will be established correctly.In this case, the segno field of both studies has already been corrected.A possible general error, which would require corrective measures of the installed base, could not be determined by the investigation.For future software versions, this rare effect will be taken into account accordingly.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIS VIBE HEMO
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstr. 1
forchheim, germany 91301
GM  91301
MDR Report Key11775886
Report Number3004977335-2021-77192
Device Sequence Number4331456
Product Code DQK
UDI-Device Identifier04056869010199
UDI-Public04056869010199
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K150493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11007641
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/19/2021
Supplement Date Received by Manufacturer06/23/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date06/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-