|
STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM CLUSTER48D; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
702-04-48D |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
04/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding intraop fracture involving a trident ii shell was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
This pi is for the intra-operative events that occurred during the revision surgery: while removing the anato stem, an intra-operative femoral fracture occurred and was treated with 2 cables from hospital inventory and implanting a restoration modular stem construct.While implanting a 48mm shell, the shell passed through the patient's acetabulum (surgeon reported this was due to poor bone quality).The surgeon went to a 50 multihole shell with 4 screws.The screws didn't provide adequate fixation in the surgeon's opinion (due to the patient's poor bone quality), so the 50mm shell was cemented in instead.Rep provided the primary and revision usage sheets and confirmed that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|