| Model Number |
PRD-SC30-125 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
04/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported oad and guide wire were received for analysis.No damage was observed on the oad.The spring tip of the guide wire was deformed but remained intact.The guide wire passed through the oad driveshaft and handle assembly with no resistance.The oad was tested, spun on all speeds, and functioned as intended.The device data log did not reveal any issues that would have contributed to the reported event.At the conclusion of the device analysis, the reported event was unable to be conclusively confirmed.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id# (b)(4).
|
| |
|
Event or Problem Description
|
|
A peripheral stealth orbital atherectomy device and viperwire guide wire were selected for treatment of a lesion in the distal posterior tibial artery.After six treatments, the oad was stuck on the guide wire, and a dissection was observed.The dissection was treated with an unplanned balloon angioplasty for three minutes.The procedure was then completed, and the patient was discharged home.
|
| |
|