BOSTON SCIENTIFIC CORPORATION EQUALIZER; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 20280
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.A 33mm x 100cm (7f) equalizer balloon was selected for use in a balloon-occluded retrograde transvenous obliteration (brto) procedure.The balloon was advanced into the renal vein and inflated to nominal pressure, volume inside the balloon was less than 19cc.After a few seconds, the balloon ruptured and saline leaked.The balloon catheter was removed from the patient and the procedure was completed with a different device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUALIZER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11776921
Report Number2134265-2021-05462
Device Sequence Number4230803
Product Code MJN
UDI-Device Identifier08714729402404
UDI-Public08714729402404
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K140273
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2020
Device Model Number20280
Device Catalogue Number20280
Device Lot Number0023346158
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2021
Initial Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
-
-