• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. TIEMANN MODEL COUDE TIP FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. TIEMANN MODEL COUDE TIP FOLEY CATHETER Back to Search Results
Model Number 0103SI16
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient stated balloon of the foley catheter popped and other one catheter fell out and would come loose while patient was moving.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be due to "pinhole in balloon or cuff wall thickness too thin".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.Sterile: unless package is opened or damaged.Do not use if package is opened or damaged.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Visually inspect the product for any imperfections or surface deterioration prior to use.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Catheters should be replaced in accordance with the cdc guideline, "guideline for prevention of catheter-associated urinary tract infection." at the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter related adverse effect, the catheter should be replaced to deflate catheter balloon: gently insert a luer slip syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the balloon of the foley catheter popped and other one catheter fell out.It would become loose while the patient was moving.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. TIEMANN MODEL COUDE TIP FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11777163
Report Number1018233-2021-02582
Device Sequence Number14452543
Product Code MJC
UDI-Device Identifier00801741017131
UDI-Public(01)00801741017131
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K040658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number0103SI16
Device Catalogue Number0103SI16
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer07/10/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-