• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII CR DEEP FLEX ISRT S1-2 15M; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII CR DEEP FLEX ISRT S1-2 15M; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED Back to Search Results
Model Number 71421534
Medical Device Problem Codes Fracture (1260); Pitted (1460); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Bone Shedding Debris (4557)
Date of Event 04/08/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a tka revision surgery the genesis ii cr deep flex insert s1-2 15m was pitted and worn, the knee capsule had poly debris in it when opened.The implant was extracted and replaced with a dished cr insert; therefore, the procedure had to be finished changing the surgical technique.Furthermore, the patient was not injured beyond the described event.No other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation but the pictures were reviewed, and the worn and pitted device was confirmed.The clinical/medical investigation concluded that, the requested clinical documentation was not available for inclusion in the medical investigation.Without the x-rays, operative reports and explant for evaluation, the clinical root cause could not be definitively concluded.The provided intra-op image of the explanted insert supports the ¿pitted and worn¿ complaint.Reportedly, the ¿capsule had poly debris in it when opened¿ and a dished cr insert was implanted; ¿ therefore, the procedure had to be finished changing the surgical technique¿.Patient impact beyond the reported poly wear, poly debris, and revision with a change in the surgical technique could not be determined.No further complications were reported.No further assessment can be rendered at this time.Should clinically relevant documentation/information and/or product evaluation findings become available, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Potential causes could include but are not limited to friction, joint tightness or lifetime of device.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII CR DEEP FLEX ISRT S1-2 15M
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11777471
Report Number1020279-2021-03745
Device Sequence Number3550728
Product Code JWH
UDI-Device Identifier03596010520333
UDI-Public03596010520333
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K041825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2016
Device Model Number71421534
Device Catalogue Number71421534
Device Lot Number06MM10484
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/08/2021
Supplement Date Received by Manufacturer08/26/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-