BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent migration occurred.The patient presented for arteriovenous fistula percutaneous transluminal angioplasty.The 60% stenosed target lesion was located in the non-tortuous and mildly calcified brachial vein.A 10x40x75 epic vascular stent was advanced to treat the lesion.However, during deployment, the stent jumped and migrated.The stent was implanted more on the cephalic vein on the distal side than originally intended.A second stent was advanced by a wire passing through the first stent, and was implanted on the lesion that was not covered by the first stent.The procedure was completed.The patient was well maintained after the procedure without any further complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11777477
Report Number2134265-2021-05384
Device Sequence Number3647402
Product Code NIO
UDI-Device Identifier08714729805014
UDI-Public08714729805014
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2025
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0026178162
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2021
Initial Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-