| Model Number |
07P9020 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Was this device service by a third party? no.An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely decreased alinity i afp result on a patient.Results provided: (b)(6) 2021 sid (b)(6) = 13 / > 1660 / > 1660 / x10 dilution = 5823 / 5800 iu/ml.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets was performed for the likely cause alinity i afp reagent lot number 23244fn00.The ticket search determined that there is normal complaint activity for this lot number.Testing was performed in-house using a retained kit of lot# 23244fn00 and all specifications were met, indicating the lot is performing acceptably.A review of the manufacturing documentation did not identify any issues associated with the customer¿s observation.Return testing was not completed as returns were not available.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency was identified for the alinity i afp reagent lot number 23244fn00.This follow up is being submitted to include d8 and h6 information previously submitted using the h10 section in their respective fields.
|
| |
|
Search Alerts/Recalls
|