• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® TIEMANN MODEL UROLOGICAL CATHETER COUDE TIP; COUDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® TIEMANN MODEL UROLOGICAL CATHETER COUDE TIP; COUDE CATHETER Back to Search Results
Model Number 010118
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bard 18 fr urological catheters were twisted and the guiding nib did not line up with the coude tip.It was stated that it was difficult to ensure the catheter tip was pointing up and there were 17 catheters from a 10 box order.The patient experienced pain and bleeding.No medical intervention was reported.
 
Event or Problem Description
It was reported that the bard 18 fr urological catheters were twisted and the guiding nib did not line up with the coude tip.It was stated that it was difficult to ensure the catheter tip was pointing up and there were 17 catheters from a 10 box order.The patient experienced pain and bleeding.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing-related.One samples was confirmed to exhibit the reported failure.The reported failure was able to be reproduced.Product was returned unopened, therefore they were not used for diagnostic or treatment purposes.The product had caused the reported failure.A potential root cause for this failure could be "machine error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review is not required as the reported event was confirmed as manufacturing related.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TIEMANN MODEL UROLOGICAL CATHETER COUDE TIP
Common Device Name
COUDE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11778603
Report Number1018233-2021-02602
Device Sequence Number3647250
Product Code KOD
UDI-Device Identifier00801741029882
UDI-Public(01)00801741029882
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number010118
Device Catalogue Number010118
Device Lot NumberNGES0042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/13/2021
Supplement Date Received by Manufacturer11/17/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date12/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-