• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA25-70/I16-30
Medical Device Problem Codes Complete Blockage (1094); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Ischemia (1942); Necrosis (1971)
Date of Event 04/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and gore (non-endologix) excluder legs to treat an abdominal aortic aneurysm (aaa); this initial implant is outside the indications of use.Approximately (20) twenty days post initial procedure, it was reported that at an unknown date the patient was found to have necrosis in the large intestine.The large intestine was removed.It's possible that the necrosis occurred because the blood flow to the area was blocked due to the embolization of both internal iliac arteries and the occlusion of the inferior mesenteric and lumbar arteries with the afx2 bifurcated stent graft.The patient was discharged after the surgery to treat the necrosis.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.Ian evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the mesenteric occlusion and ischemia is unconfirmed.This is not consistent with the reported adverse event/incident.The clinical evaluation also shows the procedure related harm identified was an additional surgical procedure (colostomy).Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the lack of records and imaging available for review.The final patient status was reported to be discharged after the additional surgical procedure.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: g1,2: contact office- name has been updated.H6 evaluation result codes; remove 3233.H6 evaluation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key11780268
Report Number2031527-2021-00188
Device Sequence Number3018525
Product Code MIH
UDI-Device Identifier00818009014870
UDI-Public(01)00818009014870(17)2342799011
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/24/2023
Device Model NumberBEA25-70/I16-30
Device Lot Number2342799-001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer05/17/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date07/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
-
-