• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S3-4RM/LL9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S3-4RM/LL9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422256
Medical Device Problem Codes Difficult to Insert (1316); Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during ukr the journey uni tibinrt s3-4rm/ll9mm did not click in the size 4 tray.No significant delay was reported and the procedure was finished with a smith and nephew back up device.No patient injuries reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, a product analysis could not be performed at this time and the reported event could not be confirmed.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY UNI TIBINRT S3-4RM/LL9MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11780669
Report Number1020279-2021-03761
Device Sequence Number3138133
Product Code HSX
UDI-Device Identifier03596010582171
UDI-Public03596010582171
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K061011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71422256
Device Catalogue Number71422256
Device Lot Number20BM20700
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/13/2021
Supplement Date Received by Manufacturer06/11/2021
Initial Report FDA Received Date05/05/2021
Supplement Report FDA Received Date06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-