| Model Number |
AIRLIFE¿ ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT M |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
03/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the airlife¿ adult manual resuscitator, 40'' (1.0m) oxygen reservoir tubing, adult mask experienced patient ambu bag was not seeming to fill with air or there was no resisting of the bag.Patient o2 saturation dropped.As an intervention, different ambu bag was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: the physical sample was received.Found broken defect on the component.According to the investigation, root cause was undetermined but the damage in the unit provided could be attributed to misuse by the customer.
|
| |
|