| Catalog Number |
7209392 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopic latarjet the dyonics's motor sounded off and stopped working midway through sawing the coracoid.The procedure was successfully completed by manually using the saw blade.A change of technique was used without delay.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that there were no indications to suggest the anticipated risk is not adequate.According to the report, the surgeon was able to change the surgical technique to complete the procedure without injury to the patient or a delay in the procedure.The patient impact was the change in procedure.Since no harm has been alleged to this patient, no further clinical/medical assessment is required at this time.A visual inspection was performed and no deficiencies were observed.Functional evaluation revealed an abnormal sound.The root cause was associated with an internal component failure.Factors that could have contributed to the reported event include rough handling of the device or wear of the device over time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|