| Model Number |
71070367 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the grey/gold driver of a ratchet wrench was frozen.It was unknown when occurred, how the procedure was finished and if there was any surgical delay.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that, the grey/gold driver of a ratchet wrench was frozen.As this was noticed after a trauma surgery during a post-operative inspection of instruments in the sterile processing department., no case involved; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, intended for use in treatment was returned for evaluation: a visual inspection was conducted and confirmed the ratchet wrench is damaged which causes it to not properly function.A review of the complaint history revealed additional complaints for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|