| Model Number |
RT200 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The rt200 is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Method: the complaint rt200 adult dual-heated breathing circuit was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information and photographs provided by the customer and our knowledge of the product.Results: visual inspection on the provided photographs revealed that the grommet of the rt200 adult dual-heated breathing circuit was missing from the inspiratory elbow.Conclusion: without the complaint device, we are unable to determine the cause of the missing grommet.All rt200 adult dual-heated breathing circuits are leak tested during production and those that fail are rejected.The subject rt200 adult dual-heated breathing circuit would have met the required specification at the time of production.The user instructions that accompany the rt200 adult dual-heated breathing circuit state: "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|
| |
|
Event or Problem Description
|
|
A healthcare facility in (b)(6) reported via the (b)(4) national medical products administration (nmpa) reporting system, that a rt200 adult dual-heated breathing circuit was found to be leaking when connected to the ventilator before patient use.It was further reported that the grommet on the inspiratory elbow was found missing.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|