Model Number W5-6-4-MVI-3
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformance's associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The investigation found the web shape to be that of an sls rather than sl as indicated on the package.This issue is tracked by quality and manufacturing.
 
Event or Problem Description
It was reported that a web device was the wrong shape as compared to the packaging label.There was no patient injury or intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Common Device Name
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
MDR Report Key11781827
Report Number2032493-2021-00172
Device Sequence Number11766493
Product Code OPR
UDI-Device Identifier00842429110898
UDI-Public(01)00842429110898(11)210218(17)240131(10)21021813Y
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K162524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberW5-6-4-MVI-3
Device Lot Number21021813Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2021
Initial Report FDA Received Date05/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-