• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LAG/COMP SCREW KIT 100/95; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LAG/COMP SCREW KIT 100/95; NAIL, FIXATION, BONE Back to Search Results
Model Number 71677100
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after placing the nail in a cephalomedular fracture reduction surgery, the cephalic fixation was carried out through the kit offered by the system and it was done step by step according to the surgical technique, but when the lag/comp screw kit 100/95 was placed, it was evident that the sliding screw and the pin moved in a different way from its initial position.For this reason the screws were removed to reposition the pin into the required position.Once this was achieved, the sliding screw was placed without novelties and the lag/comp screw kit 100/95 was placed but did not slide through the sliding screw.Therefore, an alternative s+n device (lag/comp screw kit 95/90) was used instead to completed the procedure that was carried out without any significant delays.No patient injuries reported.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device was found to be damaged along the tip likely from attempted use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.According to inspection drawing, in-process inspection during the gun drilling process includes 100% inspection of a all necessary parameters.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.The contribution of the device to the reported event could not be corroborated.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.This issue was evaluated through our internal quality process and a quality hold was initiated for this event; the device was found to be within specification.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The device, used in treatment, was not received for evaluation and the reported event could not be confirmed.The shipping information was provided and all efforts were made to locate the device, but we have no evidence that it was ever received by the complaint investigation team for evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable cause could include but are not limited to patient anatomy, alignment, procedural variance or user error.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG/COMP SCREW KIT 100/95
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11781910
Report Number1020279-2021-03807
Device Sequence Number11558243
Product Code JDS
UDI-Device Identifier03596010563323
UDI-Public03596010563323
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71677100
Device Catalogue Number71677100
Device Lot Number20ET37638
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 04/12/2021
Supplement Date Received by Manufacturer08/25/2021
07/07/2022
Initial Report FDA Received Date05/06/2021
Supplement Report FDA Received Date08/26/2021
07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient SexFemale
Patient Weight45 KG
-
-