|
It was reported that (3) fast fx strut for tsf {} med ((b)(4)), (6) fast fx strut x-short ((b)(4)), (6) fast fx strut for tsf {} sht ((b)(4)) and (1) fast fx strut for tsf {} lng ((b)(4)) are loosing the black coating on the outside making impossible to read the numbers.Some also will not allow the shoulder bolts to thread in.It is unknown whether the event happened and if there was a patient involvement.
|
|
H3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Wear potential causes could include but are not limited to friction, joint tightness or lifetime of device.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|