SMITH & NEPHEW, INC. TANDEM BIPOLAR COCR 55OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Model Number 71322055
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 04/09/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after thr surgery performed five weeks before revision surgery, the patient experienced an infection ((b)(6) bacteraemia).The revision surgery was performed on (b)(6) 2021.The following devices were explanted from the patient: cpcs ss prim so 12/14 sz 3 ((b)(4)), tandem bipolar cocr 55od 28id ((b)(4)) and cocr 12/14 fem head 28 +0 ((b)(4)).New devices were implanted.The patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the requested medical/surgical records and x-rays are not available; therefore, a thorough medical assessment cannot be rendered.The clinical root cause for the bacterial infection remains unknown.The patient impact beyond the reported infection and revision procedure itself cannot be determined.Should additional information/documentation become available, the clinical/medical task may be re-opened for further evaluation.No further medical assessment is warranted at this time.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM BIPOLAR COCR 55OD 28ID
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11781925
Report Number1020279-2021-03815
Device Sequence Number3268574
Product Code KWY
UDI-Device Identifier03596010492548
UDI-Public03596010492548
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71322055
Device Catalogue Number71322055
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/09/2021
Supplement Date Received by Manufacturer08/23/2021
Initial Report FDA Received Date05/06/2021
Supplement Report FDA Received Date08/24/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-