| Model Number |
SAM 350P |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device won't switch off.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Heartsine's investigation of the device concluded that the reported fault was attributed to corrosion on the membrane tail due to suspected storage outside of indicated conditions.Corrosion on the membrane tail did not impact the device's ability to provide shock therapy.
|
| |
|
Event or Problem Description
|
|
Device won't switch off.No patient involvement.
|
| |
|
Search Alerts/Recalls
|