| Model Number |
SAM 350P |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Difficulty inserting pad pak.When inserted the red status led was illuminated and the device would not power on.No patient involvement.
|
| |
|
Event or Problem Description
|
|
Difficulty inserting pad pak.When inserted the red status led was illuminated and the device would not power on.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Heartsine's investigation of the device could not confirm the reported fault.Throughout the investigation the device performed to specification.Information from the history log indicated a period when no log entries were recorded up to the (b)(6) 2021, with the reported fault being raised during this time on the (b)(6) 2021.This may suggest that the pad-pak had become incorrectly seated after initial installation, resulting in a poor connection, and the subsequent inability to power on the device alongside a red status led.However, given no fault found on the device and pad-pak plastics that would have inhibited the seating of the pad-pak, the investigation was unable to confirm this.
|
| |
|