• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Manufacturer Narrative
Olympus inspected the device at the service department of olympus trading (b)(4) and found followings; the device failed to boot because the power supply unit was burned out.The repair history of this device showed that it had not been repaired in the last year.The exact cause of the reported event has not yet been identified by legal.Manufacturer olympus medical systems corp.(omsc) for this device.If significant additional information is received, this report will be supplemented.
 
Event Description
A customer found that two fuses of the device were burnt during preparation for use.The device was used in combination with the olympus light source clv-180.The customer replaced these fuses, but the device did not turn on.The user replaced the device to another device and completed a gastroscopy.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Olympus medical systems corp.(omsc) assumed that the reported event was caused by burnout around the fuse socket due to deterioration.This deterioration may have been caused by long-term use.If significant additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11782243
MDR Text Key270913200
Report Number8010047-2021-05918
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCV-180
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-