| Model Number |
621-400ESR |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As per reported by the user facility: pump began to smoke and there was black charring around the outlet.No injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device involved was received for evaluation.Thermal tracing shows the event source as internal to the plug of the ac power cord (e30608) with the resultant heat/residue spreading out from the plug to the daisy chain outlet and into the pump itself (e36935) and onto the housing of e26027.The event source is seen in the cavity that was formed between the contacts of the ac power cord.The plug of the ac power cord remained shorted under load during testing.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|