• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TAN/FAN PERC DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TAN/FAN PERC DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631021
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2021
Event Type  Injury  
Event Description
It was reported that, during a trauma surgery, after using the tan/fan perc drill guide (b)(4) and tan/fan radiolucent drop (b)(4) the drill missed to target the nail.The procedure was completed with a delay greater than 30 minutes using a competitors device.The condition of the patient is unknown.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN PERC DRILL GUIDE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11783119
MDR Text Key249234022
Report Number1020279-2021-03820
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010604873
UDI-Public03596010604873
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71631021
Device Catalogue Number71631021
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/15/2021
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-