• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631022
Medical Device Problem Code Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a trauma surgery, after using the tan/ fan perc drill guide ((b)(4)) and tan/fan radilucent drop ((b)(4)) the drill missed to target the nail.The procedure was completed with a delay greater than 30 minutes using a competitors device.The condition of the patient is unknown.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN RADILUCENT DROP
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11783146
Report Number1020279-2021-03821
Device Sequence Number3268579
Product Code HSB
UDI-Device Identifier03596010604880
UDI-Public03596010604880
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631022
Device Catalogue Number71631022
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/11/2021
Supplement Date Received by Manufacturer07/15/2021
Initial Report FDA Received Date05/06/2021
Supplement Report FDA Received Date07/19/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-