| Model Number |
N/A |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2021-00080.Medical devices: tibia cemented 5 degree stemmed left size d catalog#: 42532006701 lot#: 64761364 femur cemented posterior stabilized (ps) narrow right size 6 catalog#: 42500006002 lot#: 63897746 report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tibial insert and tray intended for a left knee were implanted in a right knee.No revision procedure has been reported.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed via compatibility check.The reported products were reviewed for compatibility and it was determined that the following implants are not compatible: left-sided tibial plate and left-sided articular surface used with right-sided femoral implant in the right knee.Device history record was reviewed and no discrepancies were found.Root cause of the reported issue due to the surgeon not following the instruction.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|