• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING PS LEFT 10 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING PS LEFT 10 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2021-00080.Medical devices: tibia cemented 5 degree stemmed left size d catalog#: 42532006701 lot#: 64761364 femur cemented posterior stabilized (ps) narrow right size 6 catalog#: 42500006002 lot#: 63897746 report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a tibial insert and tray intended for a left knee were implanted in a right knee.No revision procedure has been reported.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed via compatibility check.The reported products were reviewed for compatibility and it was determined that the following implants are not compatible: left-sided tibial plate and left-sided articular surface used with right-sided femoral implant in the right knee.Device history record was reviewed and no discrepancies were found.Root cause of the reported issue due to the surgeon not following the instruction.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING PS LEFT 10 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key11784248
Report Number3007963827-2021-00101
Device Sequence Number19747616
Product Code JWH
UDI-Device Identifier00889024234277
UDI-Public(01)00889024234277(17)270630(10)64345843
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42511400510
Device Lot Number64345843
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/10/2021
Supplement Date Received by Manufacturer05/12/2021
Initial Report FDA Received Date05/06/2021
Supplement Report FDA Received Date05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-