The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # dla21, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova (b)(4) corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla21 mitroflow aortic pericardial heart valve at the time of manufacture and release.Based on the information received from the site, the patient's endocarditis was resolved.Although a moderate steno-insufficiency of the device was noted, the device was not explanted as a result of the event.Thus, it can be concluded that the functionality of the valve was confirmed to be acceptable.Given that the valve is functional, and the onset of endocarditis was delayed (approximately 7 years post-implant), the event can reasonably be concluded as not device-related, and can be attributed to the patient's specific clinical condition.As reported by the physician, the temporary catheter placement may have contributed to the event.Furthermore, blood cultures confirmed the presence of enterococcus casseliflavus.If enterococcus casseliflavus has been present on our tissue valve before sterilization, it would have been killed easily by our liquid chemical sterilant.The sterilant contains a mixture of glutaraldehyde, formaldehyde and alcohol, all of which are highly effective against staphylococci.This liquid chemical sterilization process has been validated with spore forming bacillus atrophaeus, which is the most resistant microorganism known for aldehydes.In addition, the sterile packaging solution used for shipment of crown valves is a 4% formaldehyde solution; the valves remain in this extremely antimicrobial solution until they are opened during surgery.Therefore, it is not conceivable that a staphylococci could survive on a valve exposed to our sterilant, or the packaging solution.
|
|
A patient received a mitroflow dla21 in 2013.The manufacturer was informed that the patient had developed endocarditis and was subjected to antibiotic therapy for several weeks.To date, the pet is negative.The endocarditis was diagnosed on (b)(6) 2020.The valve appeared degenerated with steno-insufficiency.The pathogen was identified as enterococcus casseliflavus.As reported, the endocarditis resolved after the antibiotic treatment provided.The patient is in excellent clinical conditions, apyretic and asymptomatic.The device was not explanted.The patient received a pacemaker in 2018.The reported possible cause is perhaps due to temporary catheter (hypothesis).The steno insufficiency is moderate, and it is not known whether it was present prior to the diagnosis of endocarditis.
|