| Model Number |
442192 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6); initial reporter phone number: (b)(6).A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus aerobic/f culture vials 2 false positive results were obtained by the laboratory personnel.The customer stated results were not reported out so there was no report of patient impact.The following information was provided by the initial reporter: "blood culture bottles flagged positive but no bacteria seen on the smear.Please verify if there was any patient impact - were any erroneous results reported prior to gram stain? no there was no erroneous results.".
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus aerobic/f culture vials 2 false positive results were obtained by the laboratory personnel.The customer stated results were not reported out so there was no report of patient impact.The following information was provided by the initial reporter: "blood culture bottles flagged positive but no bacteria seen on the smear.1.Please verify if there was any patient impact - were any erroneous results reported prior to gram stain? no there was no erroneous results.".
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: customer reported a false positive defect.Bd was unable to reproduce customer experience with the bactec product.A false positive response was not observed when retention samples were tested.Batch history records were reviewed, and all testing were within specification for product release.Users are cautioned in the package insert under limitation of the procedure: ¿a gram-stained smear from culture medium may contain small number of non-viable organisms derived from media constituents, staining reagents, immersion oil, glass slide and specimens used for inoculation.There are many factors that can influence the false positive rate, including blood volume, blood cell counts, environmental factors, and media lot to lot variations.Complaint is unconfirmed based on retention samples and batch history record review.No correctives actions were required.
|
| |
|
Search Alerts/Recalls
|