| Model Number |
2653 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor sporadically shows no values.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.No physical damage was observed during visual inspection.The sensor failed continuity testing due to a broken wire at the emitter solder joint assembly.When connected to a monitor a "replace sensor" error message was provided and the sensor led did not illuminate.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor sporadically shows no values.No patient impact or consequences were reported.
|
| |
|