H10 h3, h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.No visual issues were observed.A functional evaluation revealed the needle will deploy when trigger is initiated.The suture capture was able to pass sutures as intended and the steps were repeated several times to make sure the device was working properly.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|