Per the instructions for use (ifu), access site complications/cardiovascular injury, including perforation or dissection of vessels which may require intervention, are potential adverse events associated with the transcarotid transcatheter aortic valve replacement procedure.Vascular complications are a well recognized complication of the tavr procedure in this elderly population with multiple co-morbidities.Edwards has reviewed many reports, including screening data records and source documentation of vascular complications and has found that the root cause is typically related to a combination of vessel size, tortuosity and calcifications.Although the incidence is decreasing with smaller sheath/delivery system sizes and physician experience, there will continue to be cases in which vascular complications will occur.The thv physician training manuals instruct on procedural considerations for sheath insertion with regards to proper screening critical to reducing vascular complications.The training manual instructs the operator on proper sheath insertion and withdrawal techniques, including pre-dilating the vessel with the edwards dilators, as needed.It also notes that calcification may reduce lumen diameter and limit or prevent transfemoral passage of the devices.The ifu contraindicates patients with vessel characteristics that would preclude safe placement of sheaths such as severe obstructive calcification or severe tortuosity.Pre-procedure screening and assessment of the access artery internal diameters will enable the clinician to determine if the sapien valve can be delivered.Assessment of location and amount of circumferential calcium will aid in determining areas of reduced vessel diameters.The operators are trained to measure minimum vessel diameter taking calcium into account.The physician training manual also lists the minimum recommended vessel size for each size device.Despite the best screening tools, a small percentage of patients will have access vessels that are not capable of safely accommodating vascular access with large bore sheaths or where increased resistance is encountered during insertion of devices.In many cases, the vessel minimum luminal diameter (mld) may be borderline or below the indicated size.In addition, significant calcification and/or tortuosity, not always appreciable on imaging, could be contributing factors to the event.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The insertion difficulty and subsequent bleeding was likely related to patient's calcification catching onto the esheath during insertion.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
|
|
As found reported by the field clinical specialist, post valve deployment, bleeding was noted at the carotid insertion site.Hemostasis achieved but once the certitude sheath was removed, the bleeding persisted.Multiple sutures were needed to achieve hemostasis.The physician believed the transition of dilator/sheath got caught on some calcification and may have torn the carotid artery.There was some difficulty inserting the sheath, however the physicians felt calcification interacting with the esheath on insertion may be the cause of the injury.
|