• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problems Break (1069); Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that the contact of the video connector socket for connecting the camera head and endoscope was loose, and the endoscopic image was intermittent.This happened with multiple camera heads and flex endoscopes.There was no report of patient injury associated with the event.Olympus inspected the device at the service department of olympus (b)(4) and found that the following: the terminal of the video connector socket was bent and corroded.The video connector socket case was cracked.The printed circuit board had failure.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The exact cause of the reported event could not be conclusively determined, because the device has not been returned to olympus medical systems corp.(omsc).More than 6 years have passed since the device was delivered.The video connector socket may have become loose due to wear of the video connector socket due to repeated use for a long period of time, or due to excessive force applied when installing or removing the video connector.In addition, it is possible that the endoscopic image was not displayed due to the above causes, corrosion of the video connector socket terminal, and aging deterioration of the printed circuit board.The instruction manual provides preventive measures against reported loosening of the video connector socket.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11788631
MDR Text Key276003136
Report Number8010047-2021-05960
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-