| Model Number |
SECX-8-40-135 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a physician was attempting to use a protege rx stent to treat the common carotid artery in a patient.There were no abnormalities reported in relation to anatomy.Embolic protection was used.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).There were no issues noted when removing the device from the hoop/tray.The device prepped per the ifu without issue.The thumbscrew/lock-pin was checked for securement prior to procedure.The lock pin was removed during ifu preparation.It is reported the stent was partially pre-deployed in vivo during procedure following ifu, so it could not be used.It is unknown if intervention was required to remove the stent from the patient's vessel.There were no patient symptoms or complications associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation the device was received with the touhy-borst tight.The device was received with approximately 3mm of the stent was exposed and the stent confirmed as 40mm.The tuohy-borst was loosened and attempted to retract the stent back into the catheter.The distal struts caught up on the distal rim of the catheter and could not be retracted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|